GAS DETECTION EQUIPMENT

We are currently using the Precision MGD gas equipment, complete with HW and Chromatograph analysis.

HW ~ 100 units 1% methane

Chromatograph ~ 1% component gas

C1 – 100 units

C2 – 130 units

C3 – 90 units

iC4 – 70 units

nC4 – 60 units

*All gas equipment is tested and calibrated to industry specifications after and during each well.

MGD PRECISION GAS DETECTOR AND CHROMATOGRAPH

System Requirements

Voltage: 120 VAC 60 Cycles

Dry Compressed Air: 20 to 40 psi

 

Dimensions

Width:  19 inch

Height: 7 inches

Depth: 12 inches

Weight: 20 lbs

 

Gas Detector

Detectors:  Ceramic bead

Platinum wire

 

Meter Range:  CC detector 0-500 units

TC detector 0-5000 units

Separate zero controls and amplifiers for each detector.

Range: X1, X2, X5

Output:  0 - 5 Vdc output full scale for each detector (unranged)

Recorder output:  0 - 100 mVdc (.1 Vdc) ranged (adjustable)

CC/TC status dry contact (CC= open, TC = closed)

Adjustable audible alarm set point

Chromatograph:  Microprocessor controlled for precise cycle timing and integrator start pulses.

 

Detector:  Ceramic Bead

Range:  0-12% C1-C5

Range switch : X1, X2, X5, X10, X20

Outputs:  0-5 volts Vdc output for computer or integrator (adjustable)

Recorder output 0-10 mVdc (.01Vdc) ranged (adjustable)

Integrator start pulse

 

 

FLAME IONIZATION DETECTOR:  5 GUYS FID SYSTEM

 

FEATURES

  • Portable Modular Design 

  • Hydrocarbon, CO2, H2s, and O2 detection 

  • Web Browser UI – No Software Installation 

  • WITS, ASCII output, and integrated Log Plotting Software (TLog)

  • TCP/IP networked with embedded LTE cellular connectivity

  • Automatic shutoff for hydrogen and combustion air with power loss 

  • Remote control calibrations for unattended operation

 

SPECIFICATIONS 

  • Dual FID detectors: One for the THA and one for the GC GC Analysis Time: True 60 second throughput C1-nC5 

  • THA Response Rate < 3 second

  • GC Analysis Time: True 60 second throughput C1-nC5 

  • THA Response Rate: 

  • Database: MySQL Oracle on Solid State Drive (SSD) • 

  • Gas Trap Flow Rate: 1SCF/H to 15 SCF/H • 

  • Gas Analysis: C1, C2, C3, iC4, nC4, iC5, nC5, CO2, H2S, and O2 

  • Methane Limits of Detection: 10ppm to 100% Requirements 

  • Hydrogen: 40cc/min, 60PSI 

  • Power 110V/60Hz 75W •

  • Gas Trap 

  • Google Chrome

MASS SPECTROMETER

Gas Detection.png
Gas Detection 1.png
Gas Detection 2.png